Portable TOC Analyser for Stack Emissions POLARIS FID |
Key features
-
Rechargeable battery operations
-
Integrated hydrogen cartridge
-
Zero Air and Span Gas integrated into the chassis
-
Heated volumetric sampling system
-
Alphanumeric display and USB port for uploading methods and downloading data analysis
-
Docking station for Methane/NMHC and optional controller for heated lines
-
TÜV Approval
characteristics
-
Power supply :
-
External Power Adapter 110 - 240VAC - 24VDC @ 8,2A
-
2 batteries 9Ah (running-time 3h @ 140°C detector)
-
Size and Weight : 355(W) x 155(H) x 426(D) mm, 10 Kg (with batteries 13,4 Kg)
-
Sampling : Heated head system
-
Environmental Operating Ranges : +5°C < T < +40°C - 0% < RH < 95%
-
Sample conditions : Temperature 0 – 170°C / pressure 90 – 110kPa
-
Max detector body temperature : 190°C
-
Max temperature Sample Inlet : 180°C
-
Materials in contact with the sample : Steel AISI316 and high performance polymers
-
Sample flow : Approximately 800 ml/min
-
Sample pressure : atm ± 100 mbar
-
Measuring Range :
-
0-20/50/150/500/5000/10000 mgC/Nm 3 (other on request)
-
0-10/100/1000/5000 ppm relative to Propane
-
Detection limit (LDL) : 0,3 mg/Nm3
-
Accuracy : 1% of the F.S. or 0.4 mgC/Nm3 , whichever is greater
-
Setting Methods and Data Analysis :
-
Internal memory flash (30 days of continuous measurements)
-
Upload/download via USB
-
Flame management : Electronic, with diagnostic and ‘flame off’ visual alarm, automatic system to restart the flame
-
Hydrogen storage cartridge :
-
Metal hydrides technology (running-time about 30 hours)
-
Rechargeable via external source of hydrogen under moderate pressure
-
No restrictions on transport
|
|
|
|
|
|
|