MicroGC on-line GCX |
KEY FEATURES
-
The modular design offers the maximum flexibility in the application and it reduces the method development leadtime
-
The used TCD is up to 10 times more sensitive than the traditional high-conductivity detectors, allowing to detect the gas components at low ppm concentrations
-
The analytical modules based on micro capillary column allow analysis in few seconds
-
The rugged system for 19" rack installation allows its use in critical conditions
-
The MC2 software Suite is designed to allow non-expert users to drive the GCX and it permits instrument setup, schedule and control of monitoring activity, multi-stream analysis, advanced reports
-
The full range of sampling equipment expands the possibilities in terms of sample handling and multi-stream plants
-
The built-in LAN connectivity allows the user to control the instrument remotely
Characteristics
-
Size : 132.5mm x 483mm x 437mm - 3U Rack 19’’
-
Weight : From 10.6 Kg to 13 Kg depending on the installed modules
-
Power supply : 100 - 240 VAC
-
Operating temperature : 0°C - 50°C
-
Sampling conditions : Indoor and Outdoor with appropriate protection from atmospheric agents
-
Carrier Gas : Helium, Hydrogen, Nitrogen, Argon. Work pressure 500 kPa
-
Column temperature : Isothermal operation: 30-180 °C, 15°C above environmental temperature
-
Sample conditions : Temperature 0 - 120°C, pressure 0 - 210 kPa
-
Detector : TCD based on the Micro Electro-Mechanical System (MEMS) technology
-
Detection limits : Low ppm
-
Dynamic range : 106 ± 10%
-
Repeatability : RSD at constant temperature and pressure
(for C1 to C6 components at % level):
-
Injector type Area repeatability
-
Variable volume ≤ 1% RSD
-
Backflush, timed mode ≤ 1% RSD
-
Fixed volume ≤ 0,2% RSD
-
Backflush, fixed mode ≤ 0,5% RSD
-
Communication : TCP/IP; RS-232; Remote Digital I/O; opt.Modbus
|
|
|
|
|
|
|