Measures and technologies :
-
CO, CO2, CH4, CNHM : proprietary non dispersive dual beam infrared detectors (NDIR)
-
O2 : industrial galvanic fuel cell (ECD)
-
H2 : proprietary thermal conductivity detector (TCD)
Minimum ranges : CO, CO2, CH4, CNHM, H2, O2: 0-5%vol
Maximum ranges :
-
CO, CO2, CH4, H2, O2: 0-100%vol (intermediary ranges also available)
-
CNHM : 0-20%vol
Precision / Repeatability : ≤ ± 2% FS / ≤ 1% FS
Warm up time : 800 seconds (30 minutes to full specifications or before gas calibration)
Zero & Span Drift : ± 1% FS/week
Response time (T90) ≤ 15 s (NDIR/ ECDO2) / ≤ 30 s (TCD)
Sample Gas Conditions at analyser inlet
-
Flow rate : Nominal 1L/min (0.7 to 1.2 L/min)
-
Inlet pressure : 20 mbar mini - 500 mbar maxi
-
Outlet pressure : Atmospheric pressure
-
Temperature : Max. 50°C
-
Quality : Free of dust, tars, water, oil traces
Operation conditions
-
TAMB : 0 to 50°C
-
PAMB : 86 to 108kPa (860 to 1080 mbar)
Power supply : 220 ±44 VAC - 50Hz ± 1 Hz (power cable included)
|
|