Stack Isokinetic Sampler X1 – APIS |
-
It is available with a double head membrane pump capable of 4 m3/h or, in alternative, with a rotative pump capable of 6 m3/h. At the inlet of the line there is a dust filter easily removable from the upper panel and is also possible to install a moisture collection system.
-
The high prevalence of the pump and the electronic system for the flow management allow to automatically compensate the variations of the l loss caused by the filter clogging and to maintain the flow constant or the isokinetic flow during all the sampling period.
-
The command interface is available with an LCD display and a polyester keyboard or with a touch screen graphical display. In this way is possible to display and manage easily the great quantity of information and data.
-
It has an input for measuring the temperature in the duct with "K" thermocouple and independent sensors for the measure of the pressures (differential, static
-
and barometric) to satisfy the characteristic of "precision" required by the regulation.
-
The instrument is compatible with all the Pitot and "K" thermocouple available on the market,
-
It is available in a version with all the parameter certified by E.A laboratories as requested for the ACCREDIA accreditation.
-
The characteristic that makes it unique is the possibility to manage the acquisition of the parameters from the MEGA SYSTEM analyser equipped with the Oxygen sensor (PARAMAGNETIC) and Carbon Dioxide sensor (NDIR) for the calculation of the density of the gaseous affluent in compliance with the regulations, saving the values during the
-
sampling adjusting the isokinetic flow in real time.
-
The installation of an optional electronic card for the acquisition of signals of 4-20mA / 1-10V allowing the instrument to acquire and save in the sampling report the data coming from analyser of other manufacturer and to use some of the parameters acquir calculation of the isokinetic flow in real time.
-
On request is possible to predispose the instrument to guide the X1 - PROBE to regulate all the temperature signals saving the data in the sampling report.
-
The data measured are automatically sa be showed and saved on a USB FLASH DRIVE by the port on the upper panel.
General specification
Interface for programming management
Display touch screen (10”)
Data download interface
RS232
USB (on Flash Drive)
Operative Temperature
-10 °C ÷ +40 °C - 95% UR
Power supply
Main power [230Vac – 50Hz]
|


|
|
|
|
|
|